13. Press
ZERO
key.
14. Removethe vial from the sample chamber.
15. Place the vial (the sample) in the sample chamber
making sure that the
marks are aligned.
16. Press
TEST
key.
The result is shown in the display in mg/l Aluminium.
Notes:
1. Before use, clean the vials and the accessorieswith Hydrochloric acid (approx. 20%).
Rinsethem thoroughly with deionised water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates.The
effect of this is generally insignificant unless the water has fluoride added artificially.
In this case,the following table should be used:
Fluoride
[mg/l F]
Displayedvalue: Aluminium [mg/l Al]
0.05
0.10
0.15
0.20
0.25
0.30
0.2
0.05
0.11
0.16
0.21
0.27
0.32
0.4
0.06
0.11
0.17
0.23
0.28
0.34
0.6
0.06
0.12
0.18
0.24
0.30
0.37
0.8
0.06
0.13
0.20
0.26
0.32
0.40
1.0
0.07
0.13
0.21
0.28
0.36
0.45
1.5
0.09
0.20
0.29
0.37
0.48
---
Example:If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride
concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is
0.17 mg/l Al.
4.
Al
Al
2
O
3
Reagent
Form of reagent/Quantity
Order-No.
Set
VARIO Aluminium ECRF20
VARIO Aluminium Hexamine F 20
VARIO Aluminium ECRMasking Reagent
Powder Pack / 100
Powder Pack / 100
Liquid reagent / 25 ml
535000
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...
Page 362: ...356 MD 610_2d 11 2019...