Turbidity
10 – 1000 FAU
1.
Fill a clean vial (24 mm Ø) with
10 ml of deionised
water,
closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber and empty
the vial completely.
5.
Stir the water sample. Immediately rinse the vial with
the water sampleand fill with
10 ml water sample
.
6.
Closethe vial tightly with the capand swirl gently several
times.
7.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
8.
Press
TEST
key.
The result is shown in the display in FAU.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST