71
MD 610_2d 11/2019
12. Placethe vial in the sample chamber making sure that
the
marks are aligned.
13. Press
TEST
key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
Notes:
See page 47
*,** mg/l free Cl
*,** mg/l comb.Cl
*,** mg/l total Cl
T1 accepted
prepare T2
press TEST
Countdown
3:00
Reagent
Form of reagent/Quantity
Order-No.
Clorine Free-DPD/F10
Powder Pack / 100
530100
Chlorine Total-DPD/F10
Powder Pack / 100
530120