65
MD 610_2d 11/2019
Chlorine, total
with Vario Powder Pack
0.02 – 3.5 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, closetightly with the cap.
2. Placethe vial in the sample chamber making sure that
the
marks are aligned.
3. Press
ZERO
key.
4. Removethe vial from the sample chamber.
5. Add the contentsof
one VARIO Chlorine TOTAL-DPD/
F10 Powder Pack(blue color marking
)
straight
from the foil to the water sample.
6. Closethe vial tightly with the cap and swirl severaltimes
to mix the contents (approx. 20 seconds).
7. Placethe vial in the sample chamber making sure that
the
marks are aligned.
8. Press
TEST
key.
Wait for a
reaction period of 3 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the displayin mg/l total Chlorine.
Notes:
See page 47
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
1
2
3
4
Countdown
3:00
3
1
1