13. Placethe vial in the sample chamber making sure that
the
marks are aligned.
14. Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
Notes:
1. After use replace the bottle caps securelynoting the colour coding.
2. Store the reagent bottles in a cool, dry place ideally between 6°C and 10°C
.
3. Also see page 47
4. In sampleswith high Calcium ion contents* and/or high conductivity* can lead to
turbidity of the sample and therefore incorrect measurements.In this case,the reagent
tablets DPDNo. 1 High Calcium and DPDNo. 3 High Calcium should be used as an
alternative. (Order-No.: seereagents „Chlorine with Tablet“).
* it is not possible to give exact values,becausethe development of turbidity depends
on the nature of the sample.
T1 accepted
prepare T2
press TEST
Countdown
2:00
*,** mg/l free Cl
*,** mg/l comb.Cl
*,** mg/l total Cl
Reagent
Form of reagent/Quantity
Order-No.
Set
DPDNo. 1 buffer solution
DPDNo. 1 reagent solution
DPDNo. 3 solution
(approx. 300 tests)
3 x Liquid reagent / 15 ml
1 x Liquid reagent / 15 ml
2 x Liquid reagent / 15 ml
471056
DPDNo. 1 buffer solution
Liquid reagent / 15 ml
471010
DPDNo. 1 reagent solution
Liquid reagent / 15 ml
471020
DPDNo. 3 solution
Liquid reagent / 15 ml
471030
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...
Page 362: ...356 MD 610_2d 11 2019...