Zero accepted
prepare Test
press TEST
Countdown
5:00
12. Press
TEST
key.
Wait for a
reaction period of 5 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Molybdate /
Molybdenum.
Notes:
1. Filter turbid water samplesusing filter paper and funnel before analysis.
2. Highly buffered water samplesor extreme pH values should be adjusted to a pH of
nearly 7 with 1 mol/l Nitric acid or 1 mol/l Sodium hydroxide.
3. Concentrations above 10 mg/l Cu causestoo high test values if the reaction time of
5 minutes is increased. So it is very important to perform the test procedure as described.
4. Substanceswhich may interfere when present in concentrations at:
Aluminium
50 mg/l
Chromium
1000 mg/l
Iron
50 mg/l
Nickel
50 mg/l
Nitrite
all levels
5.
MoO
4
Mo
Na
2
MoO
4
Reagent
Form of reagent/Quantity
Order-No.
Set
VARIO Molybdenum HR1 F10
VARIO Molybdenum HR2 F10
VARIO Molybdenum HR3 F10
Powder Pack / 100
Powder Pack / 100
Powder Pack / 100
535300