152
MD 610_2d 11/2019
Iron, total (TPTZ,Note 1)
with Vario Powder Pack
0.02 – 1.8 mg/l Fe
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1.
Fill a clean vial with
10 ml deionised water
(this is the
blank).
2. Fill the second clean vial with
10 ml of the water
sample
(this is the sample).
3.
Add the contentsof
one Vario IRON TPTZF10Powder
Pack
straight from the foil into each vial.
4.
Close the vials tightly with the caps and swirl several
times to mix the contents.
5.
Press
[ ]
key.
Wait for a reaction
period of 3 minutes.
After the reaction period is finished proceed as follows
:
6.
Placethe vial (the blank) in the samplechamber making
sure that the
marks are aligned.
7.
Press
ZERO
key.
8.
Removethe vial from the sample chamber.
9.
Placethe vial (the sample) in the sample chamber mak-
ing sure that the
marks are aligned.
10. Press
TEST
key.
The result is shown in the display in mg/l Iron.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
3:00
start:
1
2
3
4
2
3
2