22
MD 610_2d 11/2019
Alkalinity-p = p-value
with Tablet
5 – 300 mg/l CaCO
3
1.
Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
,
closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5. Add
one ALKA-P-PHOTOMETER tablet
straight
from the foil to the water sample and crush the tablet
using a clean stirring rod.
6.
Closethe vial tightly with the cap and swirl severaltimes
until the tablet is dissolved.
7.
Place the vial in the sample chamber making sure
that the
marks are aligned.
8.
Press
TEST
key.
Wait for a
reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display as Alkalinity-p.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
3
5
Countdown
5:00