Notes:
1. Fill the vial with drops of the same size by holding the bottle vertically and squeeze
slowly.
2. If little or no turbidity is present at correct dose concentrations, the sample will need a
pre-concentration step in order to detect this level of polyacrylate/polymer. Carry out this
procedure as directed then test the pre-concentrated sample as above (seenext page).
3. Anomalous results occur when interferences are present as part of the product blend
or from sample contaminants. In these instancesfollow the interference removal steps
detailed below and test this treated sample as above (seenext page).
4. This test has been calibrated using polyacrylic acid 2'100 sodium salt in the range
1-30 mg/l. Other polyacrylates/polymerswill give differing responsesand therefore the
test range will vary.
Reagent
Form of reagent/Quantity
Order-No.
Set
KS255 (Polyacrylate Reagenz 1)
KS256 (Polyacrylate Reagenz 2)
Liquid reagent / 65 ml
Liquid reagent / 65 ml
56R019165
56L025565
56L025665