148
MD 610_2d 11/2019
Iron (Note 1)
with Tablet
0.02 – 1 mg/l Fe
1.
Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5. Add
one IRON LRtablet
straight from the foil to the
water sample and crushthe tablet using a clean stirring
rod.
6.
Closethe vial tightly with the cap and swirl severaltimes
until the tablet is dissolved.
7.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
8.
Press
TEST
key.
Wait for a
reaction period of 5 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Iron (Fe
2+/3+
).
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
2
0
2