179
MD 610_2d 11/2019
13. Removethe vial from the sample chamber.
14. Placethe sample in the sample chamber making sure
that the
marks are aligned.
15. Press
TEST
key.
The result is shown in the display in mg/l Molybdate /
Molybdenum.
Notes:
1. Strong alkaline or acidic water samples must be adjusted between pH 3 and pH 5 before
the reagent is added (use0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
2. Before using clean the vials with Hydrochloric acid (approx. 20%). Rinsethoroughly with
deionised water.
3.
MoO
4
Mo
Na
2
MoO
4
Reagent / Accessories
Form of reagent/Quantity
Order-No.
Set
VARIO Molybdenum 1 LRF20
VARIO Molybdenum 2 LR
Powder Pack / 100
Liquid reagent / 50 ml
535450
Mixing Cylinder
25 ml
19802650
Zero accepted
prepare Test
press TEST
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...
Page 362: ...356 MD 610_2d 11 2019...