CyA-TEST (Cyanuric acid)
with Tablet
0 – 160 mg/l CyA
1.
Fill a cleanvial (24 mm Ø) with
5 ml of the water
sample
and
5 ml deionised water (Note 1)
, close
tightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5.
Add
one CyA-TESTtablet
straight from the foil to the
prepared water sample and crush the tablet using a
clean stirring rod.
6.
Closethe vial tightly with the cap and swirl severaltimes
until the tablet is dissolved(Note 2, 3).
7.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
8.
Press
TEST
key.
The result isshown in the display in mg/l Cyanuric acid.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
6
0
1