103
MD 610_2d 11/2019
Copper, differentiated determination
with Tablet
0.05 – 5 mg/l Cu
1.
Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5.
Add
one COPPERNo. 1 tablet
straight from the foil
to the water sample and crush the tablet using a clean
stirring rod.
6.
Closethe vial tightly with the cap and swirl severaltimes
until the tablet is dissolved.
7.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
8.
Press
TEST
key.
9.
Removethe vial from the sample chamber.
10. Add
one COPPERNo. 2 tablet
straight from the foil
to the same water sample and crush the tablet using a
clean stirring rod.
11. Closethe vial tightly with the cap and swirl severaltimes
until the tablet is dissolved.
12. Placethe vial in the sample chamber making sure that
the
marks are aligned.
13. Press
TEST
key.
The result is shown in the display in:
mg/l free Copper
mg/l combined Copper
mg/l total Copper
*,** mg/l free Cu
*,** mg/l combCu
*,** mg/l total Cu
Zero accepted
prepare T1
press TEST
prepare Zero
press ZERO
T1 accepted
prepare T2
press TEST
5
0
1
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...
Page 362: ...356 MD 610_2d 11 2019...