24
MD 610_2d 11/2019
Aluminium
with Tablet
0.01 – 0.3 mg/l Al
1.
Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5. Add
one ALUMINIUM No. 1 tablet
straight from the
foil to the water sample and crush the tablet using a
clean stirring rod (dissolvethe tablet).
6. Add
one ALUMINIUM No. 2 tablet
straight from the
foil to the samewater sampleand crushthe tablet using
a clean stirring rod.
7.
Closethe vial tightly with the capand swirl gently several
times until the tablets are dissolved.
8.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
9.
Press
TEST
key.
Wait for a
reaction period of 5 minutes.
After thereaction period isfinished themeasurementstarts
automatically.
The result is shown in the display in mg/l Aluminium.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
4
0
Summary of Contents for Lovibond MD 610