3.3 Technical data
Display
Graphic Displaywith backlight
Serial Interface
Bluetooth
®
4.0 for data transfer
RJ45connector for internet updates (seechapter 2.5)
Light source
light-emitting diode – photosensor – pair arrangement
in a transparent measurement chamber
Wavelength ranges:
1 = 530 nm IF
= 5 nm
2 = 560 nm IF
= 5 nm
3 = 610 nm IF
= 6 nm
4 = 430 nm IF
= 5 nm
5 = 580 nm IF
= 5 nm
6 = 660 nm IF
= 5 nm
IF= Interference filter
Wavelength accuracy
± 1 nm
Photometric accuracy*
2% FS(T = 20°C – 25°C)
Photometric resolution
0.005 A
Measuring range of absorbance -2600 - 2600 mAbs
Protection
conforming to IP68 (1 h, 0.1 m)
Operation
Acid and solvent resistant touch-sensitive keyboard with
integral beeper as acoustic indicator.
Power supply
4 batteries (Type AA/LR 6);
lifetime: approx. 26 hours continuous useor 3500 tests
When the Bl
uetooth
®
module is activated, the battery life
will decreaseby approximately 10%.
Auto off
20 minutes after last function,
30 seconds acoustical signal before switch off
Dimensions
approx. 210 x 95 x 45 mm (unit)
approx. 395 x 295 x 106 mm (case)
Weight (unit)
approx. 450 g
Working condition
5 – 40°C at max. 30–90% relative humidity
(without condensation)
Language options
English, German, French, Spanish, Italian, Portuguese,
Polish; further languages via Internet Update
Storage capacity
500 data sets
* measuredwith standard solutions
Subject to technical modification!
To ensure maximum accuracyof test results, always use the reagent systems
supplied by the instrument manufacturer.
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...
Page 362: ...356 MD 610_2d 11 2019...