335
MD 610_2d 11/2019
Fluoride Method 170
Press[MODE], [Shift] + [4] [0] keys.
Confirm with [ ] key.
The display shows:
Press[Shift] + [3] keys.
The display shows:
1.
Fillacleanvial(24 mm Ø)with exactly
10 ml of deionised
water
, closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the marks
are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5.
Add
exactly 2 ml SPADNSreagent solution
to the
water sample.
Caution: Vial is filled up to the top!
6. Close the vial tightly with the cap and swirl gently
severaltimes to mix the contents.
<Calibration>
M170 Fluoride L
ZERO:deionised water
press ZERO
<Calibration>
1: M 191 Ca-hardness 2
2: M 191 reset 0 cali.
3: M 170 Fluoride L
Press[Shift] + [0] keys to keep the method blank.
Press[Shift] + [1] keys to erase the method blank and set
the value back to factory calibration.
The instrument goes back to mode menu automatically.