75
MD 610_2d 11/2019
13. Removethe vial from the sample chamber, empty the
vial, rinse vial and cap several times. Fill with
a few
drops of water sample
.
14. Add
one DPD No. 1 tablet
straight from the foil and
crush the tablet using a clean stirring rod.
15. Add water sample to the 10 ml mark.
16. Close the vial tightly with the cap and swirl several
times until the tablet is dissolved.
17. Placethe vial in the sample chamber making sure that
the
marks are aligned.
18. Press
TEST
key.
19. Removethe vial from the sample chamber.
20. Add
one DPDNo. 3 tablet
straight from the foil to the
same water sample and crush the tablet using a clean
stirring rod.
21. Close the vial tightly with the cap and swirl several
times until the tablet is dissolved.
22. Placethe vial in the sample chamber making sure that
the
marks are aligned.
23. Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
Chlorine dioxide in mg/l ClO
2
.
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
Notes:
Seenext page.
T1 accepted
prepare T2
press TEST
Countdown
2:00
T2 accepted
prepare T3
press TEST
*,** mg/l free Cl
*,** mg/l comb.Cl
*,** mg/l total Cl
*,** mg/l
ClO
2
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...
Page 362: ...356 MD 610_2d 11 2019...