260
MD 610_2d 11/2019
Silica HR / Silicon dioxide HR
with Vario Powder Pack
1 – 90 mg/l SiO
2
1.
Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
(Note 1), closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5.
Add the contents of
one Vario SilicaHR Molybdate
F10 Powder Pack
straight from the foil to the water
sample.
6.
Closethe vial tightly with the cap and swirl severaltimes
to mix the contents.
7.
Add the contentsof
one Vario SilicaHRAcid Rgt. F10
Powder Pack
straight from the foil to the samewater
sample (Note 2).
8.
Closethe vial tightly with the cap and swirl severaltimes
to mix the contents.
9.
Press
[ ]
key.
Wait for a
reaction period of 10 minutes
.
After the reaction period is finished proceed as follows:
10. Add the contents of
one Vario Silica Citric Acid F10
Powder Pack
straight from the foil to the water sample
(Note 3).
11. Closethe vial tightly with the cap and swirl severaltimes
to mix the contents.
12. Placethe vial in the sample chamber making sure that
the
marks are aligned.
13. Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Silica.
Countdown
2:00
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
start:
1
2
3
4
5
2
3
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...
Page 362: ...356 MD 610_2d 11 2019...