165
MD 610_2d 11/2019
10. Press
TEST
key.
Wait for a reaction
period of 15 minutes
(note 1).
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Iron.
Notes:
1. Complexed iron may be measuredby increasing the development period until no further
colour development is seen. Very strongly complexed iron may not be included in the
measured iron. In this casethe complexing agent must be destroyed by oxidation with
acid/persulphatefollowed by neutralisation to pH 6–9.
Follow the procedure on page 166.
2. For total iron (suspendedand dissolved), boil sample with acid/persulphate. Neutralise
back to pH 6–9 making back up to original volume with distilled or deionisedwater.
Follow the procedure on page 166.
Reagent / Accessories
Form of reagent/Quantity
Order-No.
KS160 – Total Hardness Buffer
Liquid reagent / 65 ml
56L016065
KS63 – Thioglycolate Reagent
Liquid reagent / 65 ml
56L006365
KP962 (Ammonium Persulphate Powder)
Pulver
56P096240
KS144 (Calcium Hardness Puffer)
Liquid reagent / 65 ml
56L014465
Spoon
0,5 g Spoon
385340
Countdown
15:00
Zero accepted
prepare Test
press TEST
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...