MD 610_2d 11/2019
2.3.5 Performing Tests
When zero calibration is complete, remove the vial from the sample chamber and perform
the tests as describedunder “Method“.
When the results have been displayed:
•
with some methods you can change between different chemical species
•
you can store and/or print out the results
•
perform further analysiswith the same zero
•
select a new method
2.3.6 Ensuring reaction periods (countdown)
To ensure compliance with reaction periods a time delay is incorporated: the countdown.
Thereare two kinds of countdowns:
• Press[ ] key.
Preparewater sample, start countdown with [ ] key and
proceed as described in the mode description.
The vial must not be placed in the sample chamber.
• Press[TEST]key.
Prepare the water sample as described in the method
description and place the vial in the sample chamber.
The display shows the countdown by pressing the [TEST]
key and the countdown is started automatically. After
the reaction period is finished the measurement starts
automatically.
Notes:
1. It is possible to finish the working countdown by pressingthe [ ] key. Readingstarts
immediately. In this case the operator is responsible for ensuring the necessaryreaction
period.
Non-compliance with reaction periods leads to incorrect test results.
2. The time remaining is displayed continuously.
The beeper indicates the last 10 seconds.
Countdown
1:59
Countdown
2:00
start:
Test
Summary of Contents for Lovibond MD 610
Page 14: ...8 MD 610_2d 11 2019...
Page 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Page 87: ...81 MD 610_2d 11 2019...
Page 95: ...89 MD 610_2d 11 2019...
Page 306: ...300 MD 610_2d 11 2019...
Page 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Page 310: ...304 MD 610_2d 11 2019...
Page 362: ...356 MD 610_2d 11 2019...