Acquiring and displaying measurements
VIBROTEST 60
Page 72 of 168
C102 853.002
© VTEST60E July 2013 4.46ff
5.1.2.2 Measurement vs. Time f(t)
Measurement window
Meas. Value 1:
Total meas. time, time interval
Unit
Signal detection
START The measurement process is started. „STOP„ is dis-
played and the channel number flashes to indicate
that the measurement is in progress.
REC
Activate the measurement recording vs. time. The
measurement process is ended manually with
„STOP“ or automatically when the set time has
elapsed.
DEL REC Delete the entire measurement without confirmation.
DEL REC deletes only the values in the buffer me-
mory and not data on the PC-card.
DISPLAY Display the measured values
during
measurement
and the recorded measurement values
after
the
measurement process is completed in the form of an
amplitude diagram. With the function button »EXIT«
you can switch back to a numeric display.
STOP
The measurement is stopped. „START“ is displayed
to indicate that the measurement has stopped.