Field Balancing
VIBROTEST 60
Page 122 of 168
C102 853.002
© VTEST60E July 2013 4.46ff
7.7
Two-plane Balancing steps with component correction
example
VIBROTEST 60 dialogue
VIBROTEST 60 steps
Activity at Rotor
Initial Run
A
B
1
2
Select Initial run with
»SELECT«.
1.
Select Initial run in the
main menu
Start the measurement with
»START«.
2.
Start Initial unbalance
measurement
The measurement is executed by
VIBROTEST 60 in the two planes
with »START«.
•
Attach the refe-
rence pickup.
•
Attach the vibration
sensors in
measuring planes 1
and 2.
•
Bring the rotor to
operational speed.
Switch over to the display of
plane 2 using the button
»1/2«.
3.
Execute the Initial
measurement
The VIBROTEST 60 executes the
measurement in the two planes.
The measured values from plane 1
are displayed.
Falls die Messwerte zu stark If the
measured values are varying strongly
the averaging function can be
activated with »AVERAGE«.