Commissioning
VIBROTEST 60
Page 138 of 168
C102 853.002
© VTEST60E July 2013 4.46ff
9
Commissioning
When the VIBROTEST 60 is switched on the main
VIBROTEST 60 menu is displayed. The dialogue language can
be changed with the softkey »SETUP« in the menu „System
Setup“.
Selectable
Measurement and
Service Functions
SELECT The highlighted item on the menu will be selected as
the next function menu.
INFO
Switch to the Information menu.
SETUP Opens the menu „System Setup„.
REPORT Opens the Reports selection list.
Using the cursor buttons the highlight bar can be moved to the
desired function. With the softkey »SELECT«, the desired
function is selected.
Independent of the position of the highlight bar, pushing the
softkey »INFO« will produce the following menu:
Version number
Fabrication number
Instrument
manufacturer
SERVICE
Opens the menu „Service“
(see section
Fehler!
Verweisquelle konnte
nicht gefunden werden.
).
Note:
The accuracy stated by the manufacturer is binding only until
the date of the „Next calibration„.
The recommended date of the next calibration is displayed. In
some cases an earlier date for the next calibration may be
sensible.
A System calibration on the date of the next recommended
calibration can be done by any authorised Brüel & Kjær Vibro
service station. At each System calibration the recommended
date is automatically changed.