Field Balancing
_________________________________________________________________________________
© VTEST60E July 2013 V4.46ff
C102 853.002
Page 123 of 168
VIBROTEST 60 dialogue
VIBROTEST 60 steps
Activity at Rotor
Stop the measurement with
»STOP«.
With the function button »1/2« the
measured values for measuring plane
2 are displayed.
The measurement at both planes is
stopped with »STOP«. This should
not be done earlier than 3 secs. after
the first values appear.
Store the measurements
with »OK«.
4.
Store Initial meas-
urement
The measurements for both planes is
stored with »OK«.
With storage the trigger level is
determined in the VIBROTEST 60.
Even the smallest variations in the
relation between reference mark and
reference pickup will cause angle
errors.
The measurement can be repeated
with »START«.
Stop the rotor.
5.
End the Initial run
With the function button »1/2« the
measured values for measuring plane
2 are displayed.
The Initial run can be repeated as
many times as required with
»SELECT«.
Test Run A
A
B
1
2
Select Test run A with
»SELECT«.
1.
Select Test run A in
the main menu