VIBROTEST 60
Field Balancing
_________________________________________________________________________________
© VTEST60E July 2013 V4.46ff
C102 853.002
Page 113 of 168
7.5
Single-plane balancing steps with polar correction
example
VIBROTEST 60 dialogue
VIBROTEST 60 steps
Activity at Rotor
Initial Run
A
1
Select Initial Run with
»SELECT«.
1.
Select Initial run in the
main menu
Start measurement with
»START«.
2.
Start Initial run
measurement
The VIBROTEST 60 starts
measurement with »START«.
•
Attach the refe-
rence pickup.
•
Attach the vibration
sensor at
measuring plane 1.
•
Bring the rotor to
operational speed.
Stop the measurement with
»STOP«.
3.
Execute initial unba-
lance measurement
The VIBROTEST 60 measures the
initial unbalance.
If the measurement is varying
strongly activate the averaging
function with »AVERAGE«.
Measurement is stopped with
»STOP«. This can be done no earlier
than 3 secs. after the first value is
displayed.