Sequence for Measurement vs. Speed
•
Activation of the measurement recording with »REC«.
This can also be activated after the measurement is started.
•
Start the measurement using »START«.
•
VIBROTEST 60 stores the measurement. Measurement is
stopped manually with »STOP« or automatically when the
range f
min
/ f
max
is completed.
•
Display the measurement as an amplitude/phase diagram
using »DISPLAY«. With »DEL REC« the last measure-
ments are deleted.
Note:
If there are any measurements stored in the memory, they must
first be deleted using »DEL REC« before further measurements
can be carried out and stored.