Acquiring and displaying measurements
VIBROTEST 60
Page 68 of 168
C102 853.002
© VTEST60E July 2013 4.46ff
5.1.2 Recording measurements in a specific speed range or time
interval
5.1.2.1 Measurement vs. speed f(n)
Measurement window
Meas. value 1:
Meas. value t 2:
Meas. value 3:
Start - Stop speed
Unit:
Signal detection
Unit:
Signal detection
Speed
START The measurement is started. „STOP„ is displayed
and the channel number flashes to indicate that the
measurement is in progress.
REC
Activate the measurement recording vs. speed or
time. The measurement process is ended manually
with „STOP“ or automatically when f
max
/ f
min
is
reached (run-up / coast-down).
DEL REC Deletes the entire measurement without a confir-
mation request. DEL REC deletes only measure-
ments in the memory but not data on the PC-Card.
DISPLAY Displays the measurement recording in an amplitude
diagram after completion.
STOP
The measurement process is stopped. The word
„START“ is displayed to indicate that the process has
stopped.
AVERAGE On / Off switch for the averaging function