61
Q-TECH™ ProgrammEr: maInTEnanCE
If for any reason the shock could not be delivered, a red background screen will appear with a
message stating “The Shock could not be delivered.”
Note:
In the event telemetry is lost, SQ-RX device commands, including Rescue
Shocks, will not be available until telemetry is reestablished.
Manual shock
Manual Shock allows the user to deliver a synchronized shock during a sinus rhythm, an atrial
rhythm or a ventricular rhythm. The shock energy level is user-configured in the 10 to 80 joule range
(Figure 52). Manual shock may also be utilized at a low energy to assess system impedance/integrity
either at implant or as warranted by patient condition. A manual shock may be administered with
the S-ICD detection zone(s) either ON or OFF.
Figure 52:
Manual shock
S-iCd System Magnet Model 4520
Model 4520 Magnet (the “magnet”) is a nonsterile accessory used to inhibit the delivery of therapy
from the SQ-RX device. Apply the magnet flat against the skin directly over the implanted SQ-RX
device for a minimum of one (1) second to suspend arrhythmia detection. Removal of the magnet
will return the SQ-RX device to normal operation. If the magnet is applied during an episode, the
episode will not be stored in the SQ-RX device memory.