33
Q-TECH™ ProgrammEr: oPEraTIon
Automatic Setup
Before the SQ-RX device can be activated, it must go through an initial Automatic Setup Process at
the time of the implant.
The Automatic Setup Process is initiated as follows:
1.
Select the Main Menu button.
2.
Select the Implant button.
3.
Select the Auto Setup button on the Navigation Bar. The Automatic Setup screen
appears (Figure 23).
4.
Select Continue if the patient’s heart rate is less than 130 bpm. For rates greater than
130 bpm, select the Cancel button and refer to the instructions for Manual Setup.
Figure 23:
Automatic Setup screen
5.
Once started the Automatic Setup will:
•
Perform the shock electrode integrity check to measure electrode impedance
(Figure 24). Normal sub-threshold impedance range is < 400 Ohms.
•
Select the best sensing configuration. The sense electrode configuration appears on
the printed report and can be viewed via the Manual Setup process.
•
Select the appropriate gain selection. The selected sense gain appears on the
Printed Report and can be viewed via the Manual Setup process.