22
Q-TECH™ ProgrammEr: oPEraTIon
wand is moved out of the telemetry communications range or if noise or interfering objects inhibit
communication. Programming commands, including Rescue Shocks, will not be available until
telemetry is reestablished.
Telemetry reconnection may occur automatically within one minute when the SQ-RX device and
wand are within telemetry range.
Note:
Whenever the programmer is in active communication with an SQ-RX
device, charging of the pulse generator in preparation for delivering a shock
(whether commanded or in response to a detected arrhythmia) is indicated
by an audible notification. The notification continues until the shock is
delivered or aborted.
Offline Behavior
When the programmer is not actively communicating with an SQ-RX device it is Offline. During
Offline sessions, programmer settings can be accessed, and stored patient sessions can be viewed
and/or printed.
Stored patient Sessions
During a patient follow-up visit, the programmer will retrieve data from the SQ-RX device’s
memory. The programmer can store up to 15 patient sessions. When the 16th session occurs,
the programmer will automatically replace the oldest stored session with the new data. A stored
session includes the following information:
•
Captured S-ECG Reports
•
Episode History (including any downloaded episodes)
•
Patient Data
•
Programmed Device Settings
To view stored patient sessions:
1.
From the Startup screen, select Stored Patient Sessions.
2.
Select the desired patient session.