Q-TECH™ ProgrammEr: gEnEral dESCrIPTIon
description
The Q-TECH Programmer (the “programmer”), Model 3200, is a component of the Cameron Health/
Boston Scientific S-ICD System, which is prescribed for patients when cardiac arrhythmia management
is warranted. The other components of the S-ICD System include the SQ-RX Pulse Generator (the “SQ-
RX device”), Q-TRAK Subcutaneous Electrode (“electrode”) and the Q-GUIDE Subcutaneous Electrode
Insertion Tool (“EIT”). The programmer communicates wirelessly with the SQ-RX device to enable
adjustment of programmable settings and data collection. The S-ICD System is designed for ease of use
and simplicity of patient management. The S-ICD System has a number of automatic functions designed
to reduce the amount of time required for implantation, initial programming and patient follow-up.
indications for Use
The S-ICD System is intended to provide defibrillation therapy for the treatment of life-threatening
ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant
ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably
terminated with anti-tachycardia pacing.
Contraindications
Unipolar pacemakers are contraindicated for use with the S-ICD System.
programmer Warnings and Cautions
general
•
External defibrillation equipment should be available for immediate use during the
implantation procedure and follow-up.
•
Use the programmer only with the external power supply packaged with the programmer.
Using other accessories may cause damage to the programmer.
•
The US power cord is for connection to 110 VAC supply mains. All other power cords are for
connection to 230 VAC supply mains. Outside North America, use the supplied power cord that
exactly matches your AC electrical outlet.
•
To avoid risk of electric shock, the programmer must only be connected to a supply mains with
protective earth.
•
Mains isolation is achieved by disconnecting the AC power cord. Do not position the programmer
or the external power supply in a manner that would make it difficult to disconnect that cord.
•
Never use a damaged external power supply or a damaged programmer.