11
Q-TECH™ ProgrammEr: oPEraTIon
To connect the wand to the programmer, slide the wand cable connector over the communication
connector port located on the rear edge of the programmer (Figure 4).
To disconnect the wand, grasp the wand cable connector and gently pull it straight off the
communication connector port.
Note:
Do not pull or yank on the cable to disconnect the wand from the
programmer. Such action could cause hidden damage to the cable. A
damaged cable might reduce wireless communication capabilities and
require a replacement wand.
Figure 4:
Connecting the wand
To obtain optimal telemetry, the wand should be placed directly over the implanted SQ-RX device.
Although the programmer may receive S-ECG telemetry at greater distances, programming should
be performed with the wand placed over the SQ-RX device.
When telemetry loss occurs, the display screen will turn yellow and a message will appear with the
text “Communication Loss” to alert the user. Reposition the wand to establish communication. The
programmer will display a screen informing the user if the SQ-RX device is found and programming
can continue.
Note:
If communication cannot be reestablished, the session should be ended and
restarted by scanning for the SQ-RX device.