53
Q-TECH™ ProgrammEr: oPEraTIon
Acquire reference S-ECg
To acquire a manual Reference S-ECG:
1.
Select the Acquire Reference S-ECG button on the Utilities menu to access the
Acquire Reference S-ECG screen.
2.
Select Continue to acquire a Reference S-ECG. The programmer will begin acquiring
the Reference S-ECG. A message will appear requesting that the patient remain still
(Figure 43). The reference S-ECG QRS template is recorded and stored in the SQ-RX
device.
3.
Select the Continue button to complete the process and return to the Utilities screen.
The Cancel button can be used at any time to end S-ECG acquisition and return to the
Utilities screen.
Figure 43:
The patient should remain still while the Reference S-ECG is acquired
Beeper Control
The SQ-RX device has an internal alert system that emits an audible beeping tone when an alert
condition occurs. Alert conditions may include Elective Replacement Indicator (ERI), End of Life
(EOL), electrode impedance out of range, prolonged charge times, failed Device Integrity Check and
irregular battery depletion. Beeper Control allows activation or suppression of the beeper.