Q-TECH™ ProgrammEr: gEnEral dESCrIPTIon
•
The programmer is not waterproof or explosion-proof and it cannot be sterilized. Do not use
it in the presence of flammable gas mixtures, including anesthetic mixtures containing air,
oxygen, or nitrous oxide.
•
Confirm that the programmer is in communication with the intended implanted SQ-RX Pulse
Generator.
•
The programmer is intended for use by healthcare professionals only. The programmer
may cause radio interference or may disrupt the operation of nearby equipment. It may be
necessary to take mitigation measures, such as reorienting or relocating the programmer or
shielding the location.
•
By design, the programmer emits radio frequencies (403.5 MHz and 2.5 GHz). This may interfere
with nearby equipment. Closely monitor medical electrical equipment in the vicinity when
using the programmer.
•
The presence of other equipment in the same frequency band (403.5 MHz for the SQ-RX Pulse
Generator and 2.5 GHz for the printer) may interfere with communication. If communication
problems occur, refer to the Troubleshooting section of this manual.
•
If the programmer is used near other medical or office equipment, all equipment should be
observed to verify normal operation. Other equipment may interfere with the programmer,
even if that other equipment complies with the International Special Committee on Radio
Interference (CISPR) emission requirements.
•
Use of non-Cameron Health/Boston Scientific-approved accessories may result in increased
emissions or decreased functionality of the programmer and may cause unintended
operational behavior of the programmer.
•
No modification of the programmer is allowed.
•
Use only the Cameron Health/Boston Scientific S-ICD programmer and appropriate software to
communicate with and program the SQ-RX Pulse Generator.
•
Do not discard the programmer in a fire, incinerate, or subject it to temperatures that exceed
100° C (212° F). This could cause the programmer to explode.
•
Do not disassemble or alter any parts of the programmer.
•
Do not immerse the programmer in liquid of any kind. If the programmer does get wet, contact
customer service for information about returning the programmer to Cameron Health/Boston
Scientific. Do not attempt to dry the programmer in an oven, microwave, or dryer because this
poses a risk of overheating or explosion.