Q-TECH™ ProgrammEr: gEnEral dESCrIPTIon
•
The S-ICD System has not been evaluated for pediatric use.
•
The S-ICD System does not provide long-term bradycardia pacing, Cardiac Resynchronization
Therapy (CRT) or Anti-Tachycardia Pacing (ATP).
Use of Other Medical Therapies/diagnostic procedures
•
External defibrillation or cardioversion may damage the S-ICD System. Avoid placing the
defibrillation paddles directly over the SQ-RX device or electrode.
•
Cardio Pulmonary Resuscitation (CPR) may temporarily interfere with sensing and may cause
delay of therapy.
•
Do not expose a patient with an implanted S-ICD System to diathermy. The interaction of
diathermy therapy with an implanted SQ-RX Pulse Generator can damage the SQ-RX Pulse
Generator and cause patient injury.
•
Do not expose the patient to MRI scanning. MRI scanning can damage the SQ-RX Pulse
Generator and cause patient injury.
•
Electrical interference or “noise” from sources such as electrosurgical and monitoring
equipment can interfere with the communication between the programmer and SQ-RX Pulse
Generator or cause inappropriate therapy. If interference occurs, move and reposition the
programmer or wand away from the source of the interference.
•
Ionizing radiation therapy, such as radioactive cobalt, linear accelerators, and betatrons
may adversely affect the S-ICD System operation. Therapeutic ionizing radiation cannot be
immediately detected; however, it can damage the electronic components of the SQ-RX Pulse
Generator. To minimize the risks of ionizing radiation:
»
Shield the SQ-RX Pulse Generator with a radiation-resistant material, regardless of the
distance between the SQ-RX Pulse Generator and the radiation beam.
»
Do not project the radiation port directly at the SQ-RX Pulse Generator.
»
Evaluate the S-ICD System operation after each radiation treatment.
•
Lithotripsy and other therapeutic forms of ultrasound can damage the SQ-RX Pulse Generator.
If required, avoid direct flow of the pulse waves near the site of the implanted SQ-RX Pulse
Generator.
•
Use caution during ablation procedures. Program the S-ICD System to Therapy Off. Keep the
current path (electrode tip to ground) as far away as possible from the implanted SQ-RX Pulse
Generator and electrode.
•
The programmer is not suitable for use in the presence of a flammable anesthetic mixture with
air or with oxygen or nitrous oxide.