Q-TECH™ ProgrammEr: maInTEnanCE
Service
There are no user-accessible or user-serviceable parts or components in the programmer. If any
service, repair, or replacement of internal compontents is needed, the programmer must be
returned to Cameron Health/Boston Scientific. Contact your Cameron Health/Boston Scientific
Representative or the customer service department for instructions and return packaging.
When requesting service, please provide information concerning the nature of the failure and the
manner in which the equipment was used when the failure occurred. The model number and serial
number should also be provided. Before returning any products to the factory, call Cameron Health/
Boston Scientific customer service or your local representative for a Return Authorization number.
Maintenance Check
Prior to each use, you should perform a visual inspection and verify the following:
•
Mechanical and functional integrity of the programmer, cables, and accessories.
•
Legibility and adherence of the programmer labels.
•
Startup screen appears a few seconds after you turn on the programmer. (The normal power-
up process verifies that the programmer has passed its internal checks and is ready for use.)
Safety Measurements
National regulations may require that the user, manufacturer, or manufacturer representative
periodically perform and document safety tests of the programmer. If such testing is required in
your country, follow the testing interval and extent of testing as regulated in your country. If you
do not know the national regulations in your country, please contact your Cameron Health/Boston
Scientific representative. If IEC/EN 62353 is a required standard in your country, but no specific
testing or interval is specified, it is recommended that you perform these safety tests using the
direct method as specified in IEC/EN 62353 at an interval of every 24 months. Test values are shown
in the Nominal Specifications table (Table 10).
programmer End of Life
The programmer and accessories are designed to provide years of service under typical use. To
dispose of, return, or exchange a programmer, contact your Cameron Health/Boston Scientific
representative or the customer service department for instructions and return packaging. Do not
dispose of the programmer in the trash or at electronics recycling facilities.