TrOUBLESHOOTiNg
64
Q-TECH™ ProgrammEr: TroUblESHooTIng
This section presents potential programmer issues and the possible solutions. Of note, restarting
the programmer can often resolve many of the issues listed below. The programmer can be
restarted by pressing and holding the power button until the system shutdown menu appears and
then choosing the “Restart” option.
Contact your Cameron Health/Boston Scientific representative or the customer service department
for additional assistance.
inability to print
If unable to print, follow the steps below:
1.
Ensure that the printer is turned On and that it contains paper and a sufficient ink
supply.
2.
Check printer feed for paper jam.
3.
Ensure, as applicable, that the wireless function is enabled on the printer or that the Cameron
Health/Boston Scientific approved wireless adapter is fully inserted into the left side of the
printer.
No printer Available
The No Printer Available screen will appear if a printer was not set up. Select the Try Again button or
refer to the Printer Selection section for instructions.
Touch Screen inactive while Connected to AC power
If the touch screen does not function while the programmer is connected AC power via the external
power supply, disconnect and reconnect the external power supply and restart the programmer.