Q-TECH™ ProgrammEr: oPEraTIon
To program the manually selected sense configuration:
1.
Select the Program button to save the sense vector and gain settings.
2.
The Acquire Reference S-ECG process is automatically enabled during the Manual
Setup Process. Select the Continue button to acquire a reference S-ECG. A
confirmation screen will appear when the captured reference S-ECG is acquired
(Figure 47).
3.
Select the Continue button.
Figure 47:
Captured Reference S-ECG confirmation
Smart Charge
Through the Smart Charge feature, the SQ-RX device charge initiation sequence adapts to the
occurrence of non-sustained ventricular arrhythmia episodes by delaying capacitor charging. This
conserves battery longevity and may prevent unnecessary shocks for non-sustained arrhythmias.
Smart Charge is enabled automatically when an untreated ventricular arrhythmia episode is
recorded. To reset the Smart Charge feature:
1.
Select the Main Menu button.
2.
Select the Utilities button on the Main Menu to display the Utilities Menu.
3.
Select the Smart Charge button. The Reset Smart Charge screen appears (Figure 48).