Q-TECH™ ProgrammEr: oPEraTIon
Additional programmer Functions
rescue Shock
The Rescue Shock button is available on the programmer display when the SQ-RX device Setup
Process is complete and an SQ-RX device is actively communicating with the programmer. During
active communication, a maximum (80 J) rescue shock can be delivered upon programmer
command.
To deliver a rescue shock:
1.
Select the red Rescue Button at the top of the programmer screen. The Rescue
Shock screen appears (Figure 49).
Figure 49:
Rescue Shock screen
2.
Select the Shock button to begin charging the SQ-RX device for a Rescue shock. A
red background screen with the word “Charging” will appear (Figure 50). Selecting
the Abort button will prevent delivery of a rescue shock and will return to the Device
Settings screen.