37
Q-TECH™ ProgrammEr: oPEraTIon
4.
Select the On/Off Therapy switch to set the desired therapy mode (Figure 29).
Figure 29:
Setting the On/Off Therapy switch
5.
Select and drag the Conditional Shock Zone (yellow) and Shock Zone (red) slider bars
to set the desired zone configuration.
•
The Shock Zone is programmable between 170 and 250 bpm in steps of 10 bpm.
•
The Conditional Shock Zone is programmable between 170 and 240 bpm in steps
of 10 bpm. Enhanced detection criteria are automatically enabled when the Condi-
tional Shock Zone is programmed.
•
When programming both the Shock Zone and Conditional Shock Zone, maintain
at least a 10 bpm difference between the two zones. If the Conditional Shock Zone
slider (yellow) is dragged over the Shock Zone slider (red), the two sliders will merge
to create a single Shock Zone.
6.
If post-shock pacing is desired, slide the Post Shock Pacing switch to the On position.
(Post-shock bradycardia pacing occurs at a non-programmable rate of 50 bpm for up
to 30 seconds. Pacing is inhibited if the intrinsic rate is greater than 50 bpm.)
7.
Select the Program button to apply the changes and program the SQ-RX device.
A screen will appear to confirm that the SQ-RX device settings were programmed
(Figure 30). If the SQ-RX device does not accept the SQ-RX device programming, a
Program Device alert screen will appear. Select the Try Again button to return to the
Device Settings screen.