36
Q-TECH™ ProgrammEr: oPEraTIon
9.
Once the Reference S-ECG acquisition process begins, a status screen appears. The
process may take up to one minute, during which the patient should remain still.
During this process, a template of the patient’s baseline QRS complex is stored in the
SQ-RX device. Select Cancel at any time to end Reference S-ECG acquisition. When
acquisition is complete, select the Continue button.
programming Therapy parameters
Once Automatic Setup has been completed, the SQ-RX device therapy parameters may be selected.
Note:
Data from clinical testing of the S-ICD System demonstrated a significant
reduction in inappropriate therapy with the activation of the Conditional
Shock Zone prior to hospital discharge.
To set the therapy parameters:
1.
Select the Main Menu button on the Navigation Bar.
2.
Select the Implant button.
3.
Select the Device Settings button on the Navigation Bar to display the Device
Settings screen (Figure 28).
Figure 28:
Device Settings screen