62
Q-TECH™ ProgrammEr: maInTEnanCE
Other behaviors of magnet application:
•
Inhibit shock therapy delivery
•
Terminate post-shock pacing therapy
•
Prohibit arrhythmia induction testing
•
Activate the SQ-RX device’s beeper with each detected QRS complex for 60 seconds
Note:
A programmer commanded Rescue Shock can override the use of the magnet
as long as the magnet was in place prior to the initiation of the programming
command. If the magnet is applied after the initial command, the Rescue
Shock will be terminated.
Note:
Magnet application does not affect wireless communication between the SQ-
RX device and the programmer.
Charging the programmer
When not in use, it is recommended that the programmer remain connected to the external power
supply. This will ensure that the internal battery is adequately charged.
Cleaning the programmer
Keep the programmer away from dust and dirt. Do not use harsh chemicals, cleaning solvents, or
strong detergents to clean the programmer and wand.
To clean the programmer and wand as needed:
1.
Turn the Programmer Off.
2.
Gently wipe the programmer screen with a soft, clean, dry cloth.
3.
Clean the programmer plastic case and the wand by wiping them with an isopropyl
alcohol-moistened cloth.
4.
Dry the programmer immediately to remove residue.