54
Q-TECH™ ProgrammEr: oPEraTIon
Once an alert condition is triggered, audible tones beep for 16 seconds every nine hours until the
alert condition is resolved. Beeper Control is automatically enabled when the SQ-RX device is taken
out of Shelf mode.
To reset the Beeper Control:
Note:
The Beeper Control display screen is enabled only when an alert condition
occurs. If an alert condition is activated, a notification screen will appear
upon connection.
1.
Select the Main Menu button to display the Main Menu.
2.
Select the Utilities button to display the Utilities Menu.
3.
Select the Beeper Control button.
4.
Select the Reset Beeper button to suspend audible beeping tones triggered by an
alert condition (Figure 44). If the alert condition is not corrected, the audible beeping
tones will be reactivated during the next automatic S-ICD System self-check.
Figure 44:
Resetting the Beeper
5.
The disable Beeper feature is enabled when ERI is reached or EOL occurs. If desired,
select disable Beeper when ERI or EOL occurs. Beeper tones are permanently
disabled.
Note:
If beeper is disabled alert conditions will not trigger audible tones.