45
Q-TECH™ ProgrammEr: oPEraTIon
6.
Each selected Episode screen also displays the programmed parameters and the
stored S-ECG data at the time of episode declaration.
7.
Select the Continue button to return to the Episodes pick list screen.
The following details are available for each episode:
Treated Episodes
Up to 128 seconds of S-ECG data is stored for each Treated Episode:
•
pre-episode S-ECg:
Up to 44 seconds
•
First shock:
Up to 24 seconds of pre-shock S-ECG and up to 12 seconds of post-shock
S-ECG
•
Subsequent shocks:
6 seconds of pre-shock and 6 seconds post-shock S-ECG
Untreated Episodes
An Untreated Episode is defined as any high-rate episode that spontaneously terminates
during the charging process, before a shock is delivered.
Up to 128 seconds of S-ECG data is stored for each Untreated Episode:
•
pre-episode S-ECg:
44 seconds of pre-episode S-ECG
•
Episode S-ECg:
Up to 84 seconds of tachycardia S-ECG data
printing reports from the programmer
printing reports
Patient reports can be printed before or after a patient session is ended. It is recommended that
a final report be printed immediately following the implant procedure. There are three patient
reports:
•
Summary Report
•
Captured S-ECG Report
•
Episode Reports
To print patient reports from either an Online or Offline session:
1.
Select the Main Menu button to display the Main Menu screen.
2.
Select the Print Reports button.