50
Q-TECH™ ProgrammEr: oPEraTIon
S-ECg Scale Settings
To adjust the real-time S-ECG amplitude and display speed scale settings:
1.
Select the S-ECG Display Settings icon located to the right of the Live S-ECG win-
dow. The S-ECG Settings screen is displayed.
2.
Select and drag the Gain or Sweep Speed Scale bars as desired (Figure 39). The S-ECG
scale will change according to the selected setting. The gain setting controls the
visual gain. The programmer defaults to Full Range for SQ-RX devices with a gain
setting of x1 and to 1/2 Range for SQ-RX devices with a gain setting of x2. The Sweep
Speed slider controls the display speed of the scrolling Live S-ECG. The nominal
sweep speed setting is 25 mm/sec.
Figure 39:
Adjusting Gain and Sweep Speed
Note:
Amplitude settings and display speed adjustments on scrolling real-time
S-ECG and Captured S-ECGs affect the display screen settings only and have
no impact on the SQ-RX device settings for sensing.