44
Q-TECH™ ProgrammEr: oPEraTIon
Viewing Stored Episodes
The SQ-RX device stores up to 25 treated and 20 untreated tachycardia episodes, which can be
viewed during a patient’s follow-up session. When the maximum number of episodes is exceeded,
the most recent episode replaces the oldest stored episode; however, the first treated episode will
not be overwritten.
Note:
If a spontaneous episode occurs during a patient’s follow-up session while
the SQ-RX device is wirelessly communicating with the programmer, the
episode will not be stored.
To view stored episodes:
1.
Select the Main Menu button.
2.
Select the Follow Up button.
3.
Select the Captured and Stored Episodes S-ECG icon from the Navigation Bar.
4.
Select the Episodes option to access the episodes pick list.
5.
Select an episode (treated or untreated) from the list (Figure 35). The selected episode
will be downloaded from the SQ-RX device and displayed.
Figure 35:
Stored Episodes screen (scrollable list)