39
Q-TECH™ ProgrammEr: oPEraTIon
defibrillation Testing
Once the SQ-RX device is implanted and Automatic Therapy is programmed On, defibrillation
testing may be conducted.
Note:
Defibrillation testing is recommended at implant to confirm the ability of the
S-ICD System to sense and convert VF.
Note:
Episode data associated with rescue shocks, manual shocks, and induction
testing is not stored in the SQ-RX device.
To induce VF and test the S-ICD System:
1.
Select the Programmer Main Menu button on the Navigation Bar to access the Main
Menu.
2.
Select the Patient Test button to setup the induction test (Figure 31).
Figure 31:
Select the Patient Test Button to access the Induction Test screen.