Q-TECH™ ProgrammEr: oPEraTIon
Note:
The SQ-RX device can deliver a maximum of five shocks per episode. At any
time, an 80 J rescue shock can be delivered by pressing the Rescue Shock
button.
7.
At any time prior to therapy delivery, the programmed energy may be aborted by
selecting the red Abort button.
8.
Select the Exit button to exit the induction process and return to the Main Menu
screen.
performing an SQ-rX Follow-up
Sensing Configuration and Automatic Setup
It is not necessary to perform Automatic Setup at each follow-up. If Automatic Setup is performed,
resulting in a vector change, sensing should be reevaluated. After the setup process is complete,
evaluate the streaming S-ECG during a pectoral exercise. Sensing performance during high rate
exercises can also be performed. Acceptable sensing will yield “S” markers synchronous to all QRS
complexes. If other markers are noted, use the Manual Setup process to evaluate other sensing
configurations.
Note:
If Manual Setup was previously used to override a sensing configuration,
careful consideration should be taken when selecting Automatic Setup.
If an update to the reference S-ECG is desired due to a change in the patient’s resting ECG, follow
the Acquire Reference S-ECG instructions.
Viewing SQ-rX device Status
Once communication is established, the programmer displays the Device Status screen with
information regarding the current episodes and battery status of the SQ-RX device.