Q-TECH™ ProgrammEr: oPEraTIon
3.
Select either standard (STD) or reverse (REV) polarity (Figure 32).
4.
Select and drag the red marker to set the desired shock energy for the first delivered
shock. The shock energy may be programmed from 10 to 80 J. A 15 J safety margin is
recommended for defibrillation testing.
Figure 32:
Setting the desired first shock energy for defibrillation testing
5.
Select the Continue button to display the next Induction Test screen.
Note:
Ensure that noise markers (“N”) are not present on the S-ECG prior to
induction. The presence of noise markers may delay detection and therapy
delivery.