29
Q-TECH™ ProgrammEr: oPEraTIon
programming the SQ-rX device at implant
This section provides the information necessary for programming the SQ-RX device during an implant
setting.
Entering Electrode information
The programmer maintains information on the implanted electrode. To record this information for a
patient’s new or replacement electrode:
1.
Select the Main Menu button.
2.
Select the Implant button.
3.
Select the Automatic Setup button in the Navigation Bar.
4.
Select Set Electrode ID button (Figure 18).
Figure 18:
Select the Set Electrode ID button to enter electrode information
Note:
ECG and heart rate information is not present on the Automatic Setup and
Electrode ID setup screens until the electrode has been connected to the SQ-
RX device.
5.
Enter the electrode model and serial number (Figure 19).